8 Connection
PIOX S72*
8.5 Temperature probe
91
UMPIOX_S72xV1-9EN, 2022-05-15
Connection of the temperature probe to the junction box
• Remove the blind plug for the connection of the temperature probe.
• Open the cable gland of the temperature probe. The compression part remains in the cap nut.
• Push the cable of the temperature probe through the cap nut and the compression part.
• Prepare the cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the reducer into the junction box.
• Screw the basic part into the reducer.
• Insert the cable into the junction box.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the temperature probe to the terminals of the junction box.
Fig. 8.21:
Preparation of the extension cable
1 – cap nut
2 – compression part
3 – basic part
4 – reducer
5 – external shield, brushed back
6 – sealing ring side
Fig. 8.22:
Junction box
1 – connection of the extension cable
2 – connection of the temperature probe
Tab. 8.18: Terminal assignment (junction box)
terminal
extension cable (KL2)
temperature probe (KL1)
1
red
red
2
gray
red/blue
3
blue
white/blue
4
white
white
110 mm
7 mm
1
2
3
4
6
5
1
2
KL2
KL1