20 Advanced functions
20.5 Remote functions
PIOX S72*
2022-05-15, UMPIOX_S72xV1-9EN
182
Reset of measured values
• Select the list item
Reset meas. val.
• Press ENTER.
The measured value output simulates a reposing application for the duration of the signal. The actual measured flow
velocity is ignored and the measured value is set to zero. All values of the physical quantity derived from the flow velocity
and the values from the calculation channels also yield zero.
The transmitter continues the measurement if the condition of the remote function is no longer met.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
• Select
No linkage
in order to deactivate the remote function.
• Press ENTER.
Reset of totalizers
• Select the list item
Reset totalizers
.
• Press ENTER.
The totalizers are set to zero. The totalizers are deactivated for the duration of the signal.
Totalizing starts at zero again, as soon as the condition for the remote function is no longer met.
When the totalizers are reset to zero with the remote function, the character
H
is displayed next to the measured value
during the measurement.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
• Select
No linkage
in order to deactivate the remote function.
• Press ENTER.
Stop of totalizers
• Select the list item
Stop totalizers
.
• Press ENTER.
The totalizers are stopped for the duration of the signal.
Totalizing will be continued with the last registered totalizer value if the condition of the remote function is no longer met.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
• Select
No linkage
in order to deactivate the remote function.
• Press ENTER.
Not for calculation
This function enables the user to define an event-based channel selection for calculation channels.
• Select the list item
Not for calc.
• Press ENTER.
If the condition for the remote function is met, this measuring channel will not be calculated in the calculation channel. The
measurement on the measuring channel continues running. The source of this remote function can be an event trigger or
a process input.
The measuring channel is available again for the calculation channel, if the condition for the remote function is no longer
met.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
• Select
No linkage
in order to deactivate the remote function.
• Press ENTER.