Output function test
The function of the output can now be tested.
• Connect a multimeter to the output.
• Select
Yes
to test the output. Select
No
to display the next menu item.
• Press ENTER.
• Select
On
to test the energized state of the output.
• Press ENTER.
The value has to be low ohmic.
• Select
Off
to test the de-energized state of the output.
• Press ENTER.
The value has to be high ohmic.
• Select
Repeat
to repeat the test,
Finish
to display the next menu item.
• Press ENTER.
.
• Select
Yes
to test the status of the output signal. Select
No
to display the next menu item.
• Press ENTER.
• Select
Passive
(idle state) or
Active
in case a event has been selected as source item.
• Press ENTER.
Depending on the selected switching condition, the value has to be either high or low ohmic.
• Select
Repeat
to repeat the test,
Finish
to display the next menu item.
• Press ENTER.