8 Connection
PIOX S72*
8.1 Transducers
69
UMPIOX_S72xV1-9EN, 2022-05-15
8.1.1.2
Transducer cable with plastic jacket and stripped ends
• Remove the blind plug for the connection of the transducer cable.
• Open the cable gland of the transducer cable. The compression part remains in the cap nut.
• Push the transducer cable through the cap nut and the compression part.
• Prepare the transducer cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the basic part into the transmitter housing.
• Insert the transducer cable into the housing.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the transducer cable to the terminals of the transmitter.
Fig. 8.2:
Connection of the transducer cable with SMB connectors to the transmitter
1 – SMB connector
2 – cable gland
Notice!
For good electromagnetic compatibility (EMC), it is important to ensure good electrical contact between the external
shield and the cap nut (and thus the housing).
Tab. 8.2:
Terminal assignment
terminal
connection
AV
transducer
(core)
AVS
transducer (internal shield)
ARS
transducer (internal shield)
AR
transducer
(core)
A
B
SNAP
DISP
DISP
MODE
FAST
MUX
NEXT
Q
ON
Q+
Q-
3x Q
OFF
LAN
USB
A+
B-
P1+
P2+
P3+
P4+
P5a
P6a
P7a
S
S
P1-
P2-
P3-
P4-
P5b
P6b
P7b
X2
X3
X_AV
KL1
1
KL12
KL7
KL8
KL9
KL10
KL4
X_AR
X_BV
X_BR
T1A
T1B
S1
T2A
T2B
T3A
T3B
S3
T4B
T4A
T1a
T1b
S1
T2a
T2b
T3a
T3b
S3
T4b
T4a
N(-)
PE
L(+)
AV
AV
S
AGN
ARS
AR
BV
BVS BGN
BRS
BR
2
1