7 Mounting
7.3 Temperature probe
PIOX S72*
2022-05-15, UMPIOX_S72xV1-9EN
66
7.3.2.3
Installation with quick release clasp
• Cut the tension strap to length (pipe circumf at least 120 mm).
• Position the temperature probe on the pipe (see Fig. 7.78).
• Place the tension strap around the temperature probe and the pipe.
• Insert the tension strap into the clasp.
• Tighten the tension strap.
• Tighten the screw of the clasp.
Caution!
The edge of the tension strap is very sharp.
Risk of injury!
→
Debur sharp edges.
→
Wear the required personal protective equipment.
→
Observe the applicable rules.
Fig. 7.80:
Quick release clasp