Annex
PIOX S72*
A Menu structure
215
UMPIOX_S72xV1-9EN, 2022-05-15
Example 2
The storage rate of all measured and diagnostic values of a certain temperature range is to be changed.
The normal storage rate of all measured and diagnostic values is 1 h. If the temperature is outside the defined range of 20...40 °C,
the storage rate should be 1 min. At the same time a record has to be carried out 10 s before and 60 s after the event.
The temperature range of 0...100 °C has to be determined via a current input of 4...20 mA.
Trigger:
Condition:
Action:
20 °C > fluid temperature > 40 °C on current input I1
I1 as trigger value outside the range 20...40 °C
store measured values within the temperature range 20...40 °C with a storage rate of 1 h
Special functions
Special functions
Inputs
Data logger
Current I1
Configuration
Enable I1
Logger channels
Yes
Start storing
→
Event-based
→
With input
Source item
→
Temperature
Storage rate
I1 Input range
→
1 h
→
4…20 mA
Storage rate (trig.)
Start of meas. range
→
1 min
→
0.0 °C
Buffer time
→П
End of meas. range
→
10 s
→
100.0 °C
Buffer time
П→
Error value
→
60 s
→
Yes
→
3.50 mA
Ringbuffer
Trigger value
Storage mode
Yes
Store totalizers
Function
Store diagnost. values
→
Out of range
Store transd. temp.
Range center
→
30.0 °C
Range width
→
20.0 °C
Glitch interval
→
1 s
Input info
Test signal
Test mea. range
1
2