11 Troubleshooting
11.1 Problems with the measurement
PIOX S72*
2022-05-15, UMPIOX_S72xV1-9EN
130
11.1.1
Error messages
display
reason
measures
NO SOUNDSP.
sound speed is invalid
check of the transducer coupling on the tube
NO TEMPERAT.
fluid temperature is invalid
check of the temperature probe connection
NO PRESSURE
fluid pressure is invalid
check of the pressure sensor connection
NO ANYOTHER
input quantity is invalid
check of all connections
NO FEEDBACK
internal intermediate result is invalid
check of the internal application data base by FLEXIM
T>TEMPBOUND
fluid temperature is greater than the
upper limit
adaptation of the internal application data base by FLEXIM
T<TEMPBOUND
fluid temperature is less than the lower
limit
adaptation of the internal application data base by FLEXIM
p>PRESSBOUND
pressure is greater than the upper limit
adaptation of the internal application data base by FLEXIM
p<PRESSBOUND
pressure is less than the lower limit
adaptation of the internal application data base by FLEXIM
c>SPEEDBOUND
sound speed is greater than the upper
limit
• field calibration according to section 10.5
• adaptation of the internal application data base by FLEXIM
c>SPEEDBOUND
sound speed is less than the lower limit
• field calibration according to section 10.5
• adaptation of the internal application data base by FLEXIM
x>OTHERBOUND
further input quantities greater than the
upper limit
adaptation of the internal application data base by FLEXIM
x<OTHERBOUND
further input quantities less than the
lower limit
adaptation of the internal application data base by FLEXIM
RANGE FAILED
application error: output of the analysis
value is not possible at the operating
point
• selection of the operating point according to the application
• adaptation of the internal application data base by FLEXIM
WN1=OUTOFBOUND
application error: output of the analysis
value is not possible at the operating
point
• selection of the operating point according to the application
• adaptation of the internal application data base by FLEXIM
INVALID COEFFS
internal calculation error
check of the internal application data base by FLEXIM
NUMERIC FAIL
internal calculation error
check of the internal application data base by FLEXIM
CAN NOT CALC
internal calculation error
check of the internal application data base by FLEXIM
K=OUT OF BOUNDS
internal calculation error
check of the internal application data base by FLEXIM
K>BOUND
analysis value is greater than the
upper limit
• field calibration according to section 10.5
• adaptation of the internal application data base by FLEXIM
K<BOUND
analysis value is lower than the upper
limit
• field calibration according to section 10.5
• adaptation of the internal application data base by FLEXIM
K NOT FOUND
internal calculation error
check of the internal application data base by FLEXIM