18 Data logger
PIOX S72*
18.1 Configuration of the data logger
165
UMPIOX_S72xV1-9EN, 2022-05-15
Event-based starting time
If it is necessary to start the storing of measured values at
a particular event, select
Event-based
as starting time.
The event is signalized via an input or event trigger. All
configured inputs and event triggers are displayed in the
scroll list.
• Select the input or the event trigger through which the
event is to be signalized.
• Press ENTER.
The storage rate is the frequency to transmit or store
measured values.
• Select in the scroll list a storage rate for storing the measured values in case the event does not occur.
• Press ENTER.
• Select
Off
if no measured values are to be stored, as long as the event does not occur.
• Press ENTER.
display
description
Immediately
The storing starts immediately.
Full 5 minutes
The storing starts in the next full 5 minutes.
Full 10 minutes
The storing starts in the next full 10 minutes.
Full 15 minutes
The storing starts in the next full 15 minutes.
Full 30 minutes
The storing starts in the next full 30 minutes.
Full hour
The storing starts in the next full 60 minutes.
Event-based
The storing starts when a defined event occurs.
Example
actual time: 09:06 am
setting:
Full 10 minutes
The storing starts at 09:10 am.
Notice!
Make sure that the time of all transmitters is synchronized.
Special functions
Data logger
Configuration
…
Start storing
→
Event-based
Storage rate w/o trig.
Storage rate (trig.)
Buffer time
→П
Buffer time
П→
see annex A
Special functions\Data logger\
Configuration\Start storing\Event-based
Special functions\Data logger\
Configuration\Start storing\
Event-based\Storage rate w/o trig.