10 Measurement
PIOX S72*
10.3 Measurement settings
113
UMPIOX_S72xV1-9EN, 2022-05-15
Measuring range
The sign of the measured value and the measuring range are determined.
• Select
Sign
if the sign of the measured values is to be considered for the output.
• Select
Absolute value
if the sign of the measured values is not to be considered for the output.
• Enter the lowest expected measured value. The unit of measurement of the source item will be displayed.
Start of meas. range
is the value assigned to the value
Output MIN
of the output range.
• Enter the highest expected measured value. The unit of measurement of the source item will be displayed.
End of meas. range
is the value assigned to the value
Output MAX
of the output range.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing
or
further information is displayed.
• Press ENTER.
20.0 mA
Other value
error value = 3.5 mA
Options\Channel A\Outputs\...\Measured values\Absolute value
Options\Channel A\Outputs\...\Start of meas. range
Options\Channel A\Outputs\...\End of meas. range
Options\Channel A\Outputs\...\Output info
Tab. 10.3: Examples for the error output (output range: 4...20 mA)
list item
output signal
I [mA]
20
4
t
d
t
0
t
1
t
I [mA]
20
4
t
d
t
0
t
1
t
Содержание PIOX S72 Series
Страница 1: ...PIOX S72 UMPIOX_S72xV1 9EN Operating instruction...
Страница 6: ...Table of contents PIOX S72 2022 05 15 UMPIOX_S72xV1 9EN 6...
Страница 196: ...21 Settings 21 9 Key lock PIOX S72 2022 05 15 UMPIOX_S72xV1 9EN 196...
Страница 225: ......