4 Basics – flow measurement
PIOX S72*
4.4 Undisturbed flow profile
19
UMPIOX_S72xV1-9EN, 2022-05-15
Tab. 4.1:
Recommended distance from disturbances
D – nominal pipe diameter at the measuring point
l – recommended distance between disturbance and transducer position
disturbance: 90° elbow
inlet: l
≥
10 D
(l
≥
2 D with disturbance correction)
outlet: l
≥
3 D
disturbance: double elbow
inlet: l
≥
10 D
(l
≥
2 D with disturbance correction)
outlet: l
≥
3 D
disturbance: double elbow out of plane (with elbow distance l
2
≥
3 D)
inlet: l
≥
10 D
(l
≥
2 D with disturbance correction)
outlet: l
≥
3 D
disturbance: double elbow out of plane (directly coupled)
inlet: l
≥
40 D
(l
≥
2 D with disturbance correction)
outlet: l
≥
3 D
l
l
l
l
l
l
l
2
l
2
l
l