15 Data transmission
FLUXUS G801
15.4 Data format
127
UMFLUXUS_G801V1-0EN, 2020-06-26
15.4
Data format
• Select the menu item
serial transmis.
• Press ENTER until
SER:kill spaces
is displayed.
• Select
on
if the space characters are not to be transmitted.
• Press ENTER.
The file size will be considerably reduced (shorter transmission time).
• Select the decimal marker to be used for floating-point numbers (point or comma).
• Press ENTER.
This setting depends on the setting of the PC operating system.
• Select the character to be used to separate columns (semicolon or tabulator).
• Press ENTER.
15.5
Data structure
First, the header is transmitted. The first 4 lines contain general information about the transmitter and the measurement.
The following lines contain the parameters for each channel.
Next, the line
\DATA
is transmitted, followed by the column titles for the corresponding channel. The measured values are
transmitted afterwards.
Special Funct.\SYSTEM settings\serial transmis.\SER:kill spaces
Special Funct.\SYSTEM settings\serial transmis.\SER:decimalpoint
Special Funct.\SYSTEM settings\serial transmis.\SER:col-separat.
Example
\DEVICE
\MODE
DATE
TIME
Par.Record
Meas.Point No.:
Pipe
Outer Diameter
Wall Thickness
Roughness
Pipe Material
Lining
Medium
Medium Temperat.
Fluid pressure
Transducer Type
Sound Path
Transd. Distance
Damping
Full-Scale Val.
Physic. Quant.
Unit Of Measure
Numb.Of Meas.Val
: G80x -XXXXXXXX
: ONLINE
: 2018-01-09
: 19:56:52
: A:F5050
: 60.3 mm
: 5.5 mm
: 0.1 mm
: Carbon Steel
: WITHOUT LINING
: Std. natural gas
: 38 C
: 1.00 bar
: xxx
: 3 NUM
: -15.6 mm
: 20 s
: 4.50 m3/h
: act. Volume flow
: [m3/h]/[m3]
: 100