16 Advanced functions
FLUXUS G801
16.10 Transducer temperature
139
UMFLUXUS_G801V1-0EN, 2020-06-26
16.10 Transducer temperature
It is possible to output the transducer temperature.
• Enter the HotCode
007043
.
• Select
yes
to display the transducer temperature during the measurement.
• Press ENTER.
• Select
yes
to store the transducer temperature.
• Press ENTER.
16.11 Activation of a binary output as alarm output
• Select the channel for which the output is to be activated in the program branch
Output Options
.
• Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
• Press ENTER until
Alarm Output
is displayed.
• Select
yes
to activate the alarm output.
• Press ENTER.
A max. of 3 independently operating alarm outputs R1, R2, R3 can be configured per channel. The alarm outputs can be
used to output information on the current measurement or to start and stop pumps, motors, etc.
16.11.1 Alarm properties
The switching condition, the holding behavior and the switching function of an alarm output can be defined.
The following 3 scroll lists are displayed:
•
func
: switching condition
•
typ
: holding behavior
•
mode
: switching function
Press
to select a scroll list in the upper line. Press
to select a list item in the lower line.
• Press ENTER to store the settings.
Notice!
The menu item
Alarm Output
will only be displayed in the program branch
Output Options
if a binary output
has been installed as alarm output (see section 13.1).
Output Options\for Channel A
Show T-transd.?
no
>YES<
Store T-transd.?
no
>YES<
R1=FUNC<typ mode
Function:
MAX