9 Measurement
FLUXUS G801
9.5 Execution of special functions
103
UMFLUXUS_G801V1-0EN, 2020-06-26
9.5
Execution of special functions
Commands that can be executed during a measurement are displayed in the upper line.
A command begins with →. If programmed, a program code has to be entered first.
• Press
until the command is displayed.
• Press ENTER.
The commands available are the following:
9.6
Determination of the flow direction
The flow direction in the pipe can be determined with the help of the displayed volumetric flow rate sign in conjunction with
the arrow on the transducers:
• The fluid flows in the direction of the arrow if the displayed volumetric flow rate is positive (e.g., 54.5 m³/h).
• The fluid flows against the direction of the arrow if the displayed volumetric flow rate is negative (e.g., -54.5 m³/h).
9.7
Stop of the measurement
The measurement is interrupted by pressing BRK if it is not protected by a program code.
Tab. 9.6:
Commands that can be executed during the measurement
command
explanation
→Adjust transd.
toggle to transducer positioning
If a program code is activated, the measurement will be continued 8 s after the last keyboard entry.
→Clear totalizer
reset the totalizers to zero
→Mux:Auto/Human
toggle the display between AutoMux and HumanMux mode
This display will not be indicated if the transmitter has only one measuring channel or only one
measuring channel is activated.
→Mux:Nextchan.
display the next channel
This display will not be indicated if the transmitter has only one measuring channel or only one
measuring channel is activated.
→Break measure
stop the measurement and return to the main menu
→Toggle FastFood
Notice!
Be careful not to stop a current measurement by inadvertently pressing BRK.
A:Mode=FastFood
54.5
m3/h
A:Mode=TransTime
54.5
m3/h