18 Settings
18.1 Dialogs and menus
FLUXUS G801
2020-06-26, UMFLUXUS_G801V1-0EN
152
18
Settings
18.1
Dialogs and menus
• Select the menu item
Dialogs/Menus
.
• Press ENTER.
18.1.1
Pipe circumference
• Select
on
if the pipe circumference is to be entered instead of the pipe diameter in the program branch
Parameter
.
• Press ENTER.
If
on
is selected for
Pipe Circumfer.
, the outer pipe diameter will still be requested in the program branch
Parameter
.
• In order to select the menu item
Pipe Circumfer.
, enter zero.
• Press ENTER.
The value displayed in
Pipe Circumfer.
is calculated on the basis of the last displayed value of the outer pipe diameter.
Example: 100 mm · π = 314.2 mm
• Enter the pipe circumference. The limits for the pipe circumference are calculated on the basis of the limits for the outer
pipe diameter.
• Press ENTER.
During the next scroll through the program branch
Parameter
, the outer pipe diameter that corresponds to the entered
pipe circumference will be displayed.
Example: 180 mm : π = 57.3 mm
Special Funct.\SYSTEM settings\Dialogs/Menus
Notice!
The settings will be stored at the end of the dialog. If the menu item is quit by pressing BRK, the changes will not be
stored.
Special Funct.\...\Dialogs/Menus\Pipe Circumfer.
Special Funct.\...\Dialogs/Menus\Pipe Circumfer.\Outer Diameter
Notice!
The edition of the pipe circumference is only temporary. When the transmitter switches back to the display of the pipe
circumference (internal recalculation), slight rounding errors may occur.
Example
entered pipe circumference: 100 mm
displayed outer pipe diameter: 31.8 mm
When the transmitter switches back to the display of the pipe circumference, 99.9 mm will be displayed.