15 Data transmission
FLUXUS G801
15.2 Terminal program
125
UMFLUXUS_G801V1-0EN, 2020-06-26
15.2.2
Offline transmission
• Start the terminal program.
• Enter the transmission parameters into the terminal program. The transmission parameters of the terminal program and
the transmitter have to be identical.
Selection of the interface for the offline transmission of data
• Select the menu item
serial transmis.
• Press ENTER until
Send Offline via
is displayed.
• Select the serial interface for the offline transmission of data.
This display will only be indicated if the transmitter has an interface RS485.
Transmitter settings
• Select the menu item
Print Meas.Val.
• Press ENTER.
The following message will be displayed if no measured values are stored.
• Press ENTER.
The following message will be displayed if the measured values are transmitted.
The progress of data transmission is displayed by a bar graph.
The following message will be displayed if an error has occurred during the serial transmission.
• Press ENTER.
• Check the connections and make sure the PC is ready to receive data.
Notice!
During the offline transmission only those data is transmitted that is stored in the transmitter.
Special Funct.\SYSTEM settings\serial transmis.
Special Funct.\SYSTEM settings\serial transmis.\Send Offline via
Special Funct.\Print Meas.Val.
NO VALUES
Print Meas.Val.
Send Header
.................
■■■■■■
.................
SERIAL ERROR
Print Meas.Val.