13 Outputs
13.2 Configuration of a frequency output as pulse output
FLUXUS G801
2020-06-26, UMFLUXUS_G801V1-0EN
116
13.2
Configuration of a frequency output as pulse output
A frequency output sends a signal with a frequency that depends on the volumetric flow rate. The frequency output can be
configured in such way that the source item can be totalized by using each period of the output signal as the increment.
Installation of a frequency output (optional)
• Select the frequency output to be installed.
• Press ENTER.
• Select
yes
to install or reconfigure the output.
• Press ENTER.
• Select
no
to uninstall the output and to return to the previous menu item in order to select another output.
• Press ENTER.
• In the scroll list, select the measuring channel to be assigned to the frequency output as the source channel.
• Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
• Select
Measuring value
(but not
Impuls
) as source item in the scroll list.
• Press ENTER.
If
Measuring value
is selected and the source item can be totalized, a request will be indicated whether the frequency
output is to be configured as a pulse output.
• Select
yes
.
• Press ENTER.
• Enter the upper limit of the frequency.
• Press ENTER.
The lower limit of the frequency and the error value will be set automatically to 0.5 Hz.
Special Funct.\SYSTEM settings\Proc. outputs\Install Output
Special Funct.\SYSTEM settings\Proc. outputs\Enable F1
Special Funct.\SYSTEM settings\Proc. outputs\...\F1 Source chan.
Special Funct.\SYSTEM settings\Proc. outputs\...\Source item\Measuring value
Special Funct.\SYSTEM settings\Proc. outputs\...\Setup as pulse ?
Special Funct.\SYSTEM settings\Proc. outputs\...\F1 Output MAX