13 Outputs
FLUXUS G801
13.1 Installation of a binary output
115
UMFLUXUS_G801V1-0EN, 2020-06-26
Assignment of a source item
One source item has to be assigned to each selected output.
• Select the physical quantity (source item) to be transmitted to the binary output by the source channel.
• Press ENTER.
The source items and their scroll lists are summarized in the following table.
Function test of the binary output
The function of the output can now be tested.
• Connect a measuring instrument to the terminals of the installed output.
• Select
Reed-Relay OFF
or
Open collect OFF
in the scroll list
Output Test
to test the de-energized state of the
output.
• Press ENTER.
• Measure the resistance at the output. The value has to be high ohmic.
• Select
yes
to repeat the test,
no
to return to the menu item
SYSTEM settings
.
• Press ENTER.
• Select
Reed-Relay ON
or
Open collect. ON
in the scroll list
Output Test
to test the energized state of the output.
• Press ENTER.
• Measure the resistance at the output. The value has to be low ohmic.
• Select
yes
to repeat the test,
no
to return to the menu item
SYSTEM settings
.
• Press ENTER.
Special Funct.\SYSTEM settings\Proc. outputs\...\Source item
Tab. 13.1: Configuration of the binary outputs
source item
list item
output
Limit
R1
limit message (
Alarm Output
R1
)
R2
limit message (
Alarm Output
R2
)
R3
limit message (
Alarm Output
R3
)
Impuls
from abs(x)
pulse without sign consideration
from x > 0
pulse for positive measured values of the volumetric flow rate
from x < 0
pulse for negative measured values for the volumetric flow rate
Special Funct.\SYSTEM settings\Proc. outputs\...\B1 Output Test\Reed-Relay OFF
Special Funct.\...\B1 Output Test\B1=ON\Again?
Special Funct.\SYSTEM settings\Proc. outputs\...\B1 Output Test\Reed-Relay ON
Special Funct.\...\B1 Output Test\B1=ON\Again?