14 Data logger
FLUXUS G801
14.6 Information regarding the data logger
123
UMFLUXUS_G801V1-0EN, 2020-06-26
14.6
Information regarding the data logger
Approx. 100 000 measurement data can be stored. Each measured value is stored with the corresponding totalizer and
optionally further measurement and diagnostic data.
According to the configuration of the data logger and the stored series of measured values, the remaining data logger
capacity will be displayed in the menu item
Special Funct.\Instrum. Inform.
• Select the menu item
Instrum. Inform.
• Press ENTER.
It is recommended to delete the old series of measured values before starting a measurement.
The type and the serial number of the transmitter are displayed in the upper line.
The remaining data logger capacity will be displayed in the lower line (here: 18 327 additional measured values can be
stored).
• Press ENTER twice to return to the main menu.
It is possible to store up to 100 series of measured values. The number of series of measured values depends on the total
number of measured values stored in the previous series of measured values.
Press
to scroll through the displays of the upper line.
If the ringbuffer is activated and has overflown at least once, the following display will be indicated:
Special Funct.\Instrum. Inform.
Fig. 14.1: Information regarding the data logger
x80x
-xxxxxxxx
Free
18327
full= 26.01/07:39
54.5
m3/h
last= 26.01/07:39
54.5
m3/h