14 Data logger
FLUXUS G801
14.1 Activation/deactivation of the data logger
119
UMFLUXUS_G801V1-0EN, 2020-06-26
14
Data logger
The transmitter has a data logger which stores the following data during the measurement:
• date
• time
• measuring point number
• pipe parameters
• fluid parameters
• transducer data
• sound path (reflection or diagonal arrangement)
• transducer distance
• damping factor
• storage rate
• physical quantity
• unit of measurement
• values of the totalizers
• diagnostic values
In order to store the data, the data logger has to be activated.
The remaining data logger capacity can be displayed.
14.1
Activation/deactivation of the data logger
• Select the channel for which the output is to be activated in the program branch
Output Options
.
• Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
• Press ENTER until
Store Meas.Data
is displayed.
• Select
yes
to activate the data logger, or
no
to deactivate it.
• Press ENTER.
14.2
Setting of the storage rate
The storage rate is the frequency to transmit or store measured values. It is set separately. If the storage rate is not set,
the previously selected storage rate will be used.
The storage interval should be at least equal to the number of activated measuring channels, min. 4 s are recommended.
• Select a storage rate or
EXTRA
.
• Press ENTER.
This display will only be indicated if
Store Meas.Data
and/or
Serial Output
are activated.
• If
EXTRA
is selected, enter the storage rate.
• Press ENTER.
Output Options\for Channel A
Output Options\...\Store Meas.Data
Output Options\...\Storage Rate
Output Options\...\Storage Rate\EXTRA