36
EN
•
Thermal disinfection is performed at a temperature
of 93°C.
The cleaning and maintenance instructions are the result of long
years of experience. Shorter cycles may be required, depending
on the application and the users‘ experiences.
The following consumables must be replaced in accordance
with the instructions given in Section 6.1 (Figure 8):
•
The device must be disconnected from the mains for
cleaning!
•
The instructions specified in the instructions for
use must be observed, particularly as regards the
recommended agents.
6.4 DISINFECTION
The different device components can be disinfected with
specific disinfectants:
•
Suction tube (instrument disinfectant),
•
Secretion container (instrument disinfectant),
•
Container lid (instrument disinfectant),
•
Device surface (surface disinfectant).
In this regard, observe the instructions already mentioned
in this section as well as the disinfectant manufacturers‘
instructions for use.
Thorough cleaning should always be performed prior to
disinfection, if such is required. As a general rule, disinfection
is advisable to prevent cross-infection and is also advisable if
the device is used in inpatient situations (e.g. hospitals, nursing
homes and/or other health care facilities), in order to limit the
risk of infections.
6.0
CLEANING, DISINFECTION, REPROCESSING
6.5
RECOMMENDED INSTRUMENT DISINFECTANTS
Disinfectant
Ingredients
(in 100 g) Manufacturer
GIGASEPT FF neu
(application concen-
trate)
Succinic acid dialdehyde
Dimethoxy tetrahydrofurane
Phosphonates, anionic tensides, non-ionic tensi-
des, scents, methyl isothiazolinone
11.9 g
3.2 g
Schülke & Mayr GmbH,
Norderstedt
Sekusept Plus
(application concen-
trate)
Glucoprotamin
25.0 g
Ecolab Deutschland GmbH,
Düsseldorf
6.6
RECOMMENDED SURFACE DISINFECTANTS
Disinfectant
Ingredients
(in 100 g) Manufacturer
Dismozon pur steril
(granulate)
Magnesium monoperoxyphthalate hexahydrate
80 g
BODE Chemie GmbH, Hamburg
Kohrsolin FF
(application solution)
Glutaral
Benzyl-C12-18-alkyldimethyl ammonium chlorides
Didecyl dimethyl ammonium chloride
5 g
3 g
3 g
BODE Chemie GmbH, Hamburg
Mikrozid sensitive
wipes
(ready to use disinfec
-
ting wipes)
Alkyl(C12-16)dimethylbenzylammoniumchlorid
(ADBAC/BKC (C12-16)); Didecyldimethylammoni
-
umchlorid (DDAC),
Alkyl(C12-14)ethylbenzylammoniumchlorid
(ADEBAC (C12-C14))
0,26 g
0,26 g
0,26 g
Schülke & Mayr GmbH,
Norderstedt
Perform
(concentrate)
Pentapotassium bis(peroxymonosulphate) bis
(sulphate)
Anionic tensides, non-ionic tensides, phosphona-
tes, soap, scents
45 g
Schülke & Mayr GmbH,
Norderstedt
Bacillol 30 Foam
(application solution)
Propane-1-ol
Propane-2-ol
Ethanol
45 g
25 g
4.7 g
BODE Chemie GmbH, Hamburg
Mikrobac forte
(concentrate)
Benzyl-C12-18-alkyldimethyl ammonium chlorides
N-(3-aminopropyl)-N-dodecylpropane-1,3-diamine
19.9 g
5 g
BODE Chemie GmbH, Hamburg
If disinfectants containing aldehyde and disinfectants containing amines are used on the same object, this can cause discolourations.
Figure 8
Bacterial filter
Suction tube
approx. 1.80 m long