DEPOLOX® Pool Compact
6. Operation
67
EN
11 Press "Enter" to save the entry.
12 Remove the conductivity sensor from the
top clip.
13 Screw the conductivity sensor into the cell
body cover and tighten the clamp connec-
tion.
14 Open the sample water inlet and outlet
again.
15 The conductivity electrode has its own
integrated temperature sensor. Variations
in the conductivity temperature measure-
ment can be adjusted via the menu
Tem.Offset.
This concludes the conductivity calibration.
6.6.5 Temperature calibration
1 Press the symbol.
2 Press the
Calibration menu.
3 Tap the measurement “Temperature.”
4 Press the “Cal. value” parameter.
5 Perform comparative temperature mea-
surement.
6 Use the keypad to enter the value to be
calibrated.
7 Press "Enter" to save the entry.
This concludes the temperature calibration.
6.7
Messages, alarms and errors
Messages, alarms and errors are displayed on the electronics module with the colored
mes-
sage symbol. Error messages can occur that can be acknowledged or that can not be acknowled-
ged. If several messages occur at the same time, the number of messages appears next to the
symbol. Press the
message symbol to display the message window. Configured alarms, mes-
sages that can be acknowledged and errors are displayed as clear text. A time-stamp shows
when the message was activated.
6.7.1 Acknowledgeable messages
6.7.2 Non-acknowledgeable messages and errors
NOTICE
They are acknowledged via the message window and the green ACK button
.
ACK
Error message
Cause
Remedy
Maximum dosing
time?
The maximum dosing time set for
a control output has been excee-
ded.
Identify the cause, e.g. chemicals
tank empty.
Check the dosing pump.
Auto tune error
Auto tune terminated with error. Chapter 3.5.6.
NOTICE
Error messages can only be rectified by eliminating the cause.
Error message
Cause
Remedy
Measured value
display flashes
Measured value is outside the
measurement range.
Check measurement range and
change, if necessary. Check dosing
or controller settings.
DI 1 flashes
Sample water flow rate recently
insufficient (delay time running).
Messwasser-Durchfluss prüfen
(ca. 33 l/h).
Schmutzfänger reinigen oder
wechseln.
Multi-Sensor falsch angeschlos-
sen oder defekt.
DI 1
Sample water flow rate insuffi-
cient for some time (delay time
elapsed).