6. Operation
DEPOLOX® Pool Compact
56
EN
Explanation of digital input settings:
•
Disabled:
Changes at the digital input
have no effect.
•
Enabled:
Changes at the digital input have
an effect if they are used in the alarm con-
figuration. Active = contact open at digital
input
•
Controller stop:
The controllers switch to
Stop (Dosing off) if the digital input is ope-
ned (e.g. Circulation off).
•
Standby:
All controllers switch to Dosing
off if the digital input is opened. The mea-
sured value display is hidden. Standby is
used when circulation is switched off and
no sample water is flowing through the
flow cell.
Example
To transfer a measurement signal, e.g. Chlo-
rine, via the mA output, the following setting is
required:
6.3.4 ”Alarm configuration” menu
Alarms 1 to 8 are configured in the “Alarm
Configuration” menu. The DEPOLOX® Pool
Compact offers the option of setting various
alarm configurations. It is not necessary to
assign a relay switching function to every
alarm. An alarm can also be used as an alarm
message without a relay. The number of avai-
lable relays that can be used as alarm relays
depends on the dosing output. The use of
relays for dosing output takes priority. For
example, if dosing contact is used as dosing
output for chlorine, relay K2 cannot be used as
an alarm relay.
When the alarms become active, they are
displayed in color via the
message symbol
and shown in the display area for alarms as an
alarm symbol. Pressing the message sym-
bol opens the message window. Here, the
alarm is displayed with time-stamp and
description.
Alarms that can be acknowledged can be con-
firmed by pressing the green Acknowledge
button
in the message window.
1 Call up the HOME menu.
2 Tap
the
symbol.
3 Tap the
Alarm configuration menu.
4 Enter the desired setting.
DI 1:
Measurement Stop (cannot be chan-
ged)
DI 2 / DI 3 / DI 4 / DI 5:
Disabled, Enabled, Controller Stop,
Standby
mA outputs 1/2/3/4
mA outputs:
off, 0 to 20 mA, 4 to 20 mA
Measurement:
Chlorine, pH, ORP (Redox),
conductivity, temperature
Signal:
Measured value, Yout
mA output
Settings
mA output
0/40 to 20 mA
Measurement
Chlorine
Signal
Measured value
NOTICE
The settings menus for the mA outputs are
only displayed with an installed mA output
card.
Digital inputs
Alarm 1/2/3/4
Designation:
A customer-specific alarm name can be
defined. This name is displayed in the
message window (main menu) when the
alarm becomes active.
Assignation:
Assignment of the alarm cause. All min.
or max. limit values, digital inputs, errors
(general) can be assigned as alarm cau-
ses. Multiple assignment is possible.
ACK