6. Operation
DEPOLOX® Pool Compact
58
EN
Hold Function:
On/Off
The hold function is used to either buf-
fer all measured values or keep them
constant during calibration. This pre-
vents the output of invalid control sig-
nals by the sensor during the calibration
process and also the output of erratic
values from measurements via mA-sig-
nal and communication interfaces. The
function is enabled when the calibration
menu is opened and disabled when the
menu is closed.
Display
Brightness:
0 to 100 %
Screensaver:
Off, 30 s, 1/5/15/30 min, 1 h
Color scheme:
Design 1 to 5
Main display:
Standard/narrow
Selection of the measurement arrange-
ment in the main display when using the
conductivity measurement
Calibrate LED:
This setting can be used for white
balance of the LED glow stick color if
color deviations occur.
Red, yellow, blue: 50 to 100 %
Time/Date - Date
Time:
00:00 to 24:00 / 00:00 to 12:00
24 h Format:
On/Off
Safety
Sample water delay time:
00:00 to 10:00 min
The sample water delay time determi-
nes the time after which dosing is deac-
tivated, e.g. in the event of sample
water Stop. While the delay time is run-
ning, “Const.” is displayed in the Auto-
matic symbol.
Feed delay time:
00:00 to 120:00 min
The feed delay time delays the start of
dosing when the device is switched on,
after switch-on, when the operating
mode has been changed, after Control-
ler STOP. The rundown of the selected
time can be interrupted by selecting the
“Start now!” button.
Max. dosing time:
00:00 to 10:00 h
The maximum dosing time determines
the length of time in which all control
outputs must reach their setpoint in the
range of the adjusted parameter ”Moni-
toring hyseresis”. The controller outputs
are switched off after this time. When
the setting is “00:00:00,” this function is
switched off.
Monitoring hysteresis:
2 to 50 %
This parameter defines a range around
the setpoint for the max. dosing time
observation.
Safety MAN. mode:
Stops dosing with sample water stop or
external stop.
Cl2 feed lockout:
On/Off
If this function is switched on, chlorine
dosing switches off automatically if the
pH value deviates too far from the pH
setpoint. The switch-off limits are defi-
ned by the pH setpoint and the parame-
ter “Switch-off range.”
Example:
Setpoint = 7.20 pH
Switch-off range = 0.40 pH
Chlorine dosing switches off at:
pH value > 7.20 + 0.40 = 7.60 pH or
pH value > 7.20 - 0.40 = 6.80 pH
pH switch-off range:
0.2 to 1.5
Defines the permissible deviation from
the pH setpoint for Cl2 feed lockout acti-
vated.
User administration
(chapter 6.3.8)
Access control:
This parameter is used to switch the
user administration on and off.
•
deactivated = no user administra-
tion; all parameters can be changed
without entering a password
•
activated = user administration ena-
bled, password protection or Level
1, 2 and 3 can be configured
Level 3:
If access control is enabled, as least
Level 3 must be used or configured.
Level 3 comprises read and write access
to all setting parameters of the device.