DEPOLOX® Pool Compact
6. Operation
69
EN
Temperature error ?
Interruption in the temperature
sensor or cable of the multi-sensor.
Check multi-sensor and cable.
mA Output 1 ?
mA Output 2 ?
mA Output 3 ?
mA Output 4 ?
Load error
The mA output cannot drive its
mA output current through the
connected current loop (500 ohm
at 20 mA max.).
Check whether the mA signal is
required at all (e.g. for plotter). If
not, switch off the output signal in
the “INPUTS/OUTPUTS” menu,
“Analog output”. Check mA signal
cable for interruption.
Hardware ?
Hardware or electronic error
Contact Service.
Data storage ?
Hardware or electronic error
Contact Service.
Cell ?
Chlorine sensor: Chlorine sensor
not screwed in. No sand cleaning.
Sensor, sensor cable or sensor
measuring module defective.
Sensor measuring module µA
measuring range exceeded.
Screw in sensor correctly.
Check sand cleaning.
Check the sensor, sensor cable or
sensor measuring module, replace
if necessary. Select higher µA mea-
surement range.
pH, F
-
and mV modules:
Sensor, sensor cable or sensor
measuring module defective.
Check the sensor, sensor cable
and sensor measuring module,
replace if necessary.
Conductivity sensor:
Sensor, sensor cable or Sensor-
measuring module defective, con-
ductivity to high or too low.
Cl2 feed lockout
activated
Deviation of measured pH value
from setpoint greater than the set
switch-off range (Cl2 feed lockout
activated).
Check pH dosing
pH tank may be empty
pH measurement may be faulty
Check/replace pH tank
Check pH measurement
Auto tune error
see chapter 3.5.6
Maximum dosing
time has been
exceeded
The maximum dosing time set for
a control output has been excee-
ded
Check measuring water, calibrat-
ion, dosing pump, chemical sto-
rage sensor, sensor cable,
measurement.
Error message
Cause
Remedy