DEPOLOX® Pool Compact
6. Operation
63
EN
6.5
Firmware update
The firmware for the device is updated using a
commercially available USB stick. The memory
size must be at least as large as the firmware
file itself. For a firmware update, the firmware
file “*.SREC” and the file “Bootload.ini” must
be copied to the USB stick. Do not use subdi-
rectories.
1 De-energize the electronics module.
2 Remove the housing cover of the electro-
nics module.
3 Insert the USB stick into the USB port (A).
4 Switch on mains voltage.
5 Using an insulated screwdriver, briefly
press the Update button (B) on the
motherboard.
6 The USB stick must now be removed.
7 Close the housing cover.
8 Switch the device on.
9 All settings must now be entered again
(chapter 5.15).
10 Calibrate the sensors.
Fig. 4
Section, circuit board
A USB socket
B
Update button
6.5.1 LED glow stick color signaling
The color of the LED glow stick switches bet-
ween white, yellow and red depending on the
operating state (chapterl 6.3.6).
LED white
•
All OK.
•
The device is working trouble-free.
•
No active errors or currently no message
in the message system.
LED yellow
•
Alarm that is configured as “unlatched”
has been activated. As soon as the cause
was rectified and the alarm is inactive, the
yellow color signal changes again.
•
Fault message became active (only with
enabled tank monitoring and if the Min
message from the Cl2 or pH tank is pre-
sent.)
LED red
•
Error message present.
•
Alarm that is configured with acknowled-
gment has been activated.
Name link 1...3:
Freely configurable menu name for the
devices of the IP link 1...3. A maximum
of three IP links is possible. This name is
displayed in the menu.
Menu IP link
NOTICE
If several DEPOLOX® Pool Compact decives
are connected via the IP link, when entering
the IP link, the text “/main.shtml” must be
entered after the IP address.
Example: 192.168.200.12/main.shtml
NOTICE
A firmware update can be downloaded free
of charge from the homepage of Evoqua
Water Technologies GmbH.
DANGER
Risk of injury or death!
External voltages may still be connected
even if the operating voltage is switched off.
WARNING
Risk of injury or damage to the device!
Only authorized and qualified electricians
are permitted to connect the DEPOLOX®
Pool Compact electrically and to open the
housing.
Pool 1
Spa
Outdoor pool
Sauna plunge pool
X
NOTICE
The update takes approximately 1 to 2 minu-
tes. The LED flashes green while the update
is in progress. The update is complete when
the green LED no longer flashes.
A
B