5. Installation
DEPOLOX® Pool Compact
44
EN
5.11.1 Installing calibration aids
Two calibration clips are installed in the hou-
sing cover. They are pushed into the side of
the basic housing at the back. The clip with the
plastic insert for the sensor is pushed into the
top catch (A). The second clip is for holding
calibration solution or buffer solution in a bag
or beaker. For solution in a bag, fit the clip in
the top position of the bottom catch (B). For
solution in a beaker, select the position below
this (C).
Fig. 9
Rear side of flow cell
A Top holding clip for sensor mounting
B
Position of holding clip for bag
C
Position of holding clip for beaker
5.12 Electrical installation
A
B
C
DANGER
Risk of injury or death!
External voltages may still be connected
even if the operating voltage is switched off.
Disconnect all power sources before ope-
ning the electronics module.
WARNING
Risk of injury or damage to the device!
•
Only authorized and qualified electrici-
ans are permitted to install the DEPO-
LOX® Pool Compact and open the
housing.
•
The electronics module may only be put
into operation when the housing is clo-
sed, and must be connected to protec-
tive earth.
•
Modifications to the device other than
those described in this instruction
manual are not permissible.
•
The electronics module may only be
wired in de-energized state.
•
Connect the electronics module in
accordance with the wiring diagram
(chapter 9.) and applicable local and
national regulations.
•
High temperatures at the terminals of
the relays and the mains supply!
At high ambient temperatures, high
temperatures can occur at the terminals,
and the connected cables must be desi-
gned to withstand such temperatures.
Ambient temperature <30°C (<86°F):
cable temperature resistant up to at
least 60°C
Ambient temperature <40°C (<104°F):
cable temperature resistant up to at
least 70°C
Ambient temperature >40°C (>104°F):
cable temperature resistant up to at
least 80°C