Installation
Teqwave F/I Modbus TCP
18
Hauser
6.2.2
Mounting the transmitter
1
2
A0035459
7
Mounting the transmitter
1
DIN rail holder
2
DIN rail according to DIN EN 60715 TH 35
Mount the transmitter on the DIN rail using the DIN rail holder.
6.3
Post-mounting check
Perform the following checks once you have installed the measuring device:
Device condition and specifications
Notes
Is the device damaged (visual inspection)?
Does the device meet the specifications, such as:
• Ambient temperature
• Nominal pressure
• Measuring range → 68
Installation
Notes
Is there a sufficient distance between the sensor and the next pipe elbow? → 16
Teqwave I: Is the active sensor area fully immersed in the liquid? → 16
If installing Teqwave I in a pipe:
Is the black dot on the process connection aligned with the flow direction? → 68
If installing Teqwave I in a horizontal position:
Align the dot on the process connection so that it points either up or down? → 68
Process environment/process conditions
Notes
Are the specifications for the inlet run observed?
Inlet run ≥ 10 × DN
Are the specifications for the outlet run observed?
Outlet run ≥ 5 × DN
Is the measuring device protected against precipitation and direct sunlight?