Teqwave F/I Modbus TCP
Diagnostics and troubleshooting
Hauser
59
Bit
Hexadecimal
Diagnostic message
Description
Measures
9
0x00000200
Process disturbance
The value measured for the
dispersion is greater than 1.
1.
Remove air bubbles
and/or particles.
2.
Take recommended
mounting position
into consideration
→ 16.
3.
If the error persists,
contact Endress
+Hauser Service.
10
0x00000400
Sensor configuration failed:
sensor data do not match
the transmitter software
Missing calibration.
Contact Hauser
Service.
11
0x00000800
12
0x00001000
16
0x00010000
System error
Communication error
between internal
processors.
1.
Restart the
transmitter.
2.
If the error persists,
contact Endress
+Hauser Service.
Modbus communication errors
Code
Name
Description
0x00
-
No error
0x01
Illegal Function
The selected function code is not permitted in
the transmitter. Always use Read Input Registers
(0x04) to read measured values.
0x02
Illegal Data Address.
The selected start address is not permitted in the
transmitter. Only use the start addresses listed
to read measured values.
12.5 Diagnostics information via dispersion indicator
The dispersion indicates a disturbance in the liquid caused by the dispersion of gas bubbles
or particles. This causes a widening in the group and phase speed of sound.
The measuring device displays a standardized factor. If this value is less than 1, this
indicates that the speed of sound determined for the undisturbed fluid can still be
determined with the specified measured error. On the other hand, the density measured
error can already be greater than the measured error specified under reference conditions
at values over 0.25.
Therefore, if the speed of sound and temperature are used for concentration
measurement, the value of 1 should not be exceeded. If the density is used, the value of
0.25 should not be exceeded.
If the measured dispersion is greater than the configured switch point, the measuring
device does not display the concentration any more. The configured switch point can be
viewed as follows:
Navigation using transmitter with touch screen: "Settings" → "Application parameters" →
"Diagnosis" → "Process disturbance" → "Switch point"
Navigation using the Viewer: menu "Teqwave Transmitter" → "Display filter" → "Process
disturbance" → "Switch point"