Commissioning
Teqwave F/I Modbus TCP
38
Hauser
10.4.5 Configuring the measuring range
The
Measurement range
menu contains parameters for configuring the measuring range.
• The settings made in this menu also apply for the functions of the analog output,
where they define the minimum (0 V/4 mA and maximum (10 V/20 mA).
• A valid measuring range must be configured for the selected concentration app in
order to ensure the function works correctly. In addition, the calibration range is
also specified for measured variables that are relevant for the calibration of the
concentration app (temperature and concentration). The measuring device takes
this calibration range from the concentration app file. It cannot be modified.
Navigation using transmitter with touch screen
1. Settings menu → "Application parameters" → "Measuring range" → "Select measured
variable" → "Measuring range Max"
2. Settings menu → "Application parameters" → "Measuring range" → "Select measured
variable" → "Measuring range Min"
3. Settings menu → "Application parameters" → "Measuring range" → "Select measured
variable" → "Decimal places"
Navigation using the Viewer
1. Menu "Teqwave Transmitter" → "Measuring range" → "Select measured variable" →
"Measuring range Max"
2. Menu "Teqwave Transmitter" → "Measuring range" → "Select measured variable" →
"Measuring range Min"
3. Menu "Teqwave Transmitter" → "View settings" → "Select measured variable" →
"Decimal places"
The operating tool communicates the settings to the transmitter as soon as the
Apply
button is clicked.
Parameter overview with brief description
Parameter
Procedure
Selection/input Factory setting
Measuring range
Max
Enter the upper limit for the selected measured variable or 20 mA current/
10 V voltage, while paying attention to the measuring range limits → 68
and the concentration app data sheet.
Positive decimal • Concentration 1-2: 100.00
• Temperature: 120.0
• Speed of sound: 2000.00
• Density: 1500.00
Measuring range
Min
Enter the lower limit for the selected measured variable or 4 mA current/0 V
voltage, while paying attention to the measuring range limits → 68 and
the concentration app data sheet.
Positive decimal • Concentration 1-2: 0.00
• Temperature: 0.0
• Speed of sound: 500.00
• Density: 500.00
Decimal places
Select the number of decimal places for the value of the measuring range.
• 0
• 1
• 1 ±0.5
• 2
• 2 ±0.5
• 3
• 3 ±0.5
• 4
• Concentration 1-2: 2
• Temperature: 1
• Speed of sound: 2
• Density: 1 ±0.5
10.4.6 Configuring the relay output
The
Relay output
menu contains all the parameters for configuring the relay output.
Navigation using transmitter with touch screen
1. Settings menu → "Application parameters" → "Relay output" → "Settings" → "Select
measured value" → "Output mode"