Teqwave F/I Modbus TCP
Commissioning
Hauser
37
Parameter overview with brief description
Parameter
Procedure
Selection/input
Factory setting
Analog channel 1-4
Select the measured variable or
special function to be output at the
analog output.
• If
No analog output
is
selected, the analog
interface is set to 0 V or 2
mA. The presentation on
the display continues to run
normally.
• If
Suspend measurement
is selected, the measuring
device stops and all the
measured values and the
system status suspend. If
necessary, the
Suspend
time
function can be used
to specify a delay for
switching the function on
and off.
• Concentration 1-2
• Temperature
• Speed of sound
• Density
• Dispersion
• Deactivated analog output
• Suspend measurement
Concentration 1
Current/Voltage
Select the type of signal of the
analog interface.
• Current (4 to 20 mA)
• Voltage (0 to 10 V)
Current (4 to 20 mA)
Analog output settings
Behavior of the interface if the
measuring range is breached (value
over-range/under-range).
• If
0V/2mA exceeding
limits
is selected, an error
value is displayed.
• If
Min/Max exceeding
limits
is selected, the value
output is limited to the
specified limit value
→ 38.
• 0V/2mA exceeding limits
• Min/Max exceeding limits
0 V/2 mA exceeding limits
Suspend time (s)
Enter the suspend time in seconds
until the measured values are
recorded if the Suspend
measurement function is active.
If a
suspend time (s)
is
entered and the
Suspend
measurement
function is
selected simultaneously in the
Analog channel 1-4
parameter and activation is
via the digital input, the
Suspend measurement
is
delayed by the configured
time.
Positive integer 0 to 10
000 s
0 s
Test signal
Transmitter
Enter the current or voltage value
for simulation. Output the test
signal with
Activate / Deactivate
.
Viewer
A new window appears as soon as
the
Test signal
button is pressed.
Enter the current or voltage value
for simulation. Output the test
signal with the
Test signal
button.
A new window appears as
soon as the
Test signal
button is pressed. Enter the
current or voltage value for
simulation. Output the test
signal with the
Test signal
button.
Floating-point number with sign
0.0