10.4.1 Selecting the concentration app
A concentration app is enabled via the menu
Select concentration app
(transmitter with a
touch screen) or
Manage concentration app
(Viewer).
Navigation using transmitter with touch screen
Settings menu → "Select concentration app"
Navigation using the Viewer
Menu "Teqwave Transmitter" → "Manage concentration apps"
Parameter overview with brief description
Parameter/function
Procedure
Selection/input
Factory setting
Select concentration app
(transmitter)
Manage concentration apps
(Viewer)
Transmitter
Tap the desired concentration app
to select it. If the selected
concentration app is inactive, tap
the
Activate application
button to
activate the concentration app.
Viewer
Select the desired concentration
app by selecting "concentration
apps" from the dropdown menu. If
the selected concentration app is
inactive, tap the
Activate
button to
activate the app.
Concentration app 1-n
Concentration app 1
10.4.2 Setting the concentration offset
Manual offsets are configured for the calculated concentrations via the menu
Select
concentration app
(transmitter with a touch screen) or
Concentration offset
(Viewer).
This allows users to adapt the measuring technology to different process conditions (e.g.
higher line pressure).
Navigation using transmitter with touch screen
Settings menu → Select concentration app → "Application settings" → "Concentration 1-2
Offset"
The
Application settings
function can only be accessed if the concentration app was
activated beforehand with the
Activate application
function.
Navigation using the Viewer
Menu "Teqwave Transmitter" → "Concentration offset" → "Offset"
The operating tool communicates the settings to the transmitter as soon as the
Apply
button is clicked.
Parameter overview with brief description
Parameter/function
Procedure
Selection/input
Factory
setting
Concentration 1-2 Offset
(transmitter)
Concentration 1-2/ [
unit
]
(Viewer)
Enter the value for the concentration offset.
Access to the offset setting is only possible for the active concentration
app. However, the settings are kept if the user switches concentration
app.
Floating-point number
with sign
0.00
10.4.3 Configuring the measuring unit
All the measured values are configured via the
Measuring unit
menu (transmitter with
touch screen) or via the
View settings
menu (Viewer).
• The measured value is automatically converted if the unit is changed.
• The unit for the speed of sound is m/s and cannot be changed.