Fieldgate FXA42
Technical data
Hauser
115
13.3.6 Installation height as per IEC61010-1 Ed.3
Generally up to 2
000 m (6
560 ft) above sea level
13.3.7 Degree of protection
IP20, NEMA1
13.3.8 Shock resistance
DIN EN 60068-2-27: ±15 g; 11 ms
13.3.9 Vibration resistance
EN 60068-2-64 / IEC60068-2-64: 20..2000 Hz 0.01 g
2
/Hz
13.3.10 Electromagnetic compatibility
• Interference immunity: as per IEC 61326, industrial environment
• Interference emissions: as per IEC 61326, Class B
13.4 Certificates and approvals
13.4.1 CE mark
The device meets the legal requirements of the applicable EC directives. These are listed in
the corresponding EC Declaration of Conformity along with the standards applied.
13.4.2 RoHS
The measuring system complies with the substance restrictions of the Restriction on
Hazardous Substances Directive 2011/65/EU (RoHS 2).
13.4.3 Other standards and guidelines
Other standards and guidelines that have been considered in the design and development
of the device:
• EN 60529
Degrees of protection provided by enclosures (IP code)
• EN 61010-1
Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory
Use
• IEC/EN 61326
"Emission in accordance with Class A requirements". Electromagnetic compatibility (EMC
requirements).
13.5 Telecommunications approval
13.5.1 Europe
This device meets the requirements of the Radio Equipment Directive (RED) 2014/53/EU.
13.5.2 USA and Canada
This device complies with Part 15 of the FCC rules.