EK280
Application Manual
29
b)
Setting the high frequency pulse generator's input mode:
Press the key
until "
Md.I2
" is displayed.
Press the ENTER
button. The set value flashes.
Repeatedly press one of the
or
arrow keys until the text "
HF pulses
" flashes.
Press the ENTER key to confirm the value setting. Input can be canceled using the ESC key.
c)
Setting the high frequency pulse generator's cp value:
Press the key
until "
cp.I2
" is displayed.
Press the ENTER
button. The set value flashes.
Move the cursor with the
or
arrow keys to the digits and update them with the
or
arrow
keys to the cp value of the pulse generator connected to input 2.
Press the ENTER key to confirm the value setting. Input can be canceled using the ESC key.