EK280
Application Manual
23
4.4.2
Parameterization
Parameterization is here described with reference to the EK280 keyboard and the above mentioned example.
Alternatively, parametrization can be carried out using the
enSuite
program. The paths which are subsequently
described can be found in
enSuite
under "
Parameterize online
" > "
Parameters
".
a) Set input mode of high-frequency pulse generator (input 2):
Press the key
until "
Md.I2
" is displayed.
Press the ENTER
button. The set value flashes.
Repeatedly press one of the
or
arrow keys until the text "
HF pulses
" flashes.
Press the ENTER key to confirm the value setting. Input can be canceled using the ESC key.
b) Set pc value of high-frequency pulse generator (input 2):
Press the key
until "
cp.I2
" is displayed.
Press the ENTER
button. The set value flashes.
Move the cursor with the
or
arrow keys to the digits and update them with the
or
arrow
keys to the cp value of the pulse generator connected to input 2.
Press the ENTER key to confirm the value setting. Input can be canceled using the ESC key.
c) Set parameter for low-frequency pulse generator (input 1):
Set the parameters for the low-frequency pulse generator on input 1 as described in Chapter 4.1
(page 18).
d) Setting automatic conversion:
Move the cursor to tab "
Serv.
" and to the comparison function mode "
MdCmp
" via the following path:
Serv.
Volume
Actual volume
Volume comparison
MdCmp
Press the ENTER
button. The actual parameterized value flashes.
Repeatedly press one of the
or
arrow keys until the text "
HF-LF sw.ovr
" flashes.
Press the ENTER button to confirm the set value. It is possible to abort the entry by pressing the
ESC button.
Furthermore, set the source for the fail-safe input (NF input) to "
Sc1Vm
" to "
I1=Count. inp
"
And the priority source (HF transmitter) "
Sc2Vm
" to "
I2=comp.inp
."