11. Test parameter
Full-electronic HV-Tester F1-1x Typs
11.2
Setting parameters insulation resistance test IS
On this menu the settings for the
insulation resistance test are made. The
options F1-1x E01 (DC) and E02 (ISO) are required for the IS test
.
IS tests are made using DC, with the test object isolated from the mains.
The insulation resistance between live parts and the housing is measured
and evaluated. The parameters are largely identical to the parameters for
the high voltage test (chapter 11.1).
Depending on which device variant you have, some of the parameters
given in the following may not be available or it may not be possible edit
some parameters. The values entered are changed in the parameter set
current open. The parameters in the following relate to the standard device
with 3000 V/6000 V/ 100 mA.
In case of devices with options fitted, the parameters may vary.
On this
topic see chapter 14 Options.
11.2.1 TEST VOLTAGE
Voltage range 200 ... 3000 / 300 ... 6000 V DC“
200 ... 3000 V DC
- low voltage range.
300 ... 6000 V DC
- high voltage range.
Test voltage „3000 V / 6000 V“
- enter the test voltage with which the IS
testing is to be undertaken. For example:
low
voltage range
200 ... 3000 V DC
high
voltage range
300 ... 6000 V DC
Fig. 11.6: IS test parameters - Test voltage