82
11. Test parameter
Full-electronic HV-Tester F1-1x Typs
The test voltage is then increased linearly from the start voltage to the test
voltage in one time unit.
on -
voltage range from 100 V AC/ from 200V DC (option)
off
- no ramp function, the full test voltage is applied.
Start voltage
Test voltage
U
(V)
t (s)
Start voltage (kV)
- defines the voltage value from which the high volt-
age ramp starts.
Minimum voltage „on/off“
- activates the test voltage monitoring. The
test voltage must not drop below the defined minimum voltage during
the test time. If this is the case, the device indicates U>Umin. (Only
with integrated "voltage readback" see also Chap. 14.2 - Option E04).
Aus (off)
- no minimum voltage monitoring.
Ein (on)
- minimum voltage monitoring active.
Minimum voltage (kV)
- the voltage must not drop below the pre-defined
minimum voltage during the test time.