102
12. UNIVERSAL TEST STEPS
Full-electronic HV-Tester F1-1x Typs
12.5 Output step (Option)
See also Chapt. 6.2.3.3 "Digital outputs OPT1 - 6".
The output step is only visible, if the option OPT ( digital outputs E06) is
integrated in the tester and also activated. Option E06 enhanced the tester
with 6 additional outputs.
The state of the additional outputs OPT1 ... 6, is switched in the test plan.
For this, insert the output step at an appropriate place in the test plan.
Activate (ON) or deactivate (OFF) at one or more outputs. All outputs are
transistor outputs (positive (+) switching). Specification 24 V / 50 mA.
Fig. 12.7: Output step
Recommendation
for testers in common with ´the functional test unit.
See also Chapt. 3.4.
Safety tests (HV/ ISO):
Use Output OPT1 for switch on the safety tests
(test voltage, sense lines).
Functional test
. Use Output OPT2 for switch on functional test (mains
voltage).
The remaining outputs can be used e.g. for the control of external switch-
ing devices.