40
Full-electronic HV-Tester F1-1x Typs
4.
Operating modes
4.3
Current tripping modes
MA_CT_COMPLEX
Tripping based on the complex current
MA_CT_COMPLEX_DI
Tripping based on the complex current and Delta-I
MA_CT_REAL
Tripping based on the real current
MA_CT_REAL_DI
Tripping based on the real current and Delta-I
MA_CT_PEAK
Tripping based on the peak value of the current
MA_CT_PEAK_DI
Tripping based on the peak value of the current and Delta-I
Delta-I trigger current:
XXX.X 0-50 mA/ms
range 1 mA
0-500 mA/ms
range 10 mA
0-5000
mA/ms
range 100 mA