95
11. Test parameter
Full-electronic HV-Tester F1-1x Typs
11.3.3 Test time
Continuous test „on/off
“ - switches between continuous burning and
burning with test time.
off
- test time active. The burning voltage is applied for the duration of
the test time.
on
- continuous test active. The burning voltage is applied continuously.
The burn function must be interrupted manually using the red „Reset“
button S2/H2. The continuous test can last max. 999.9 seconds.
Test time (s)
- the test time can be set from 000.1 to 999.9s. The test time
is irrelevant in the case of „Continuous test on“.
Fig. 11.13:BURN parameters - Test time