11. Test parameter
Full-electronic HV-Tester F1-1x Typs
11.4.3 Tripping
Min. current (A)
- the stated current must flow during the function test. If
the current drops below this value , a fault I < Imin is indicated.
Max. current (A)
- the stated current must flow as a maximum during
the function test. If the current is above this value , a fault I > Imax is
indicated.
Min. power (W)
- the test object must draw as a minimum the power en-
tered during the function test. If the power drops below this value , a
fault P < Pmin is indicated.
Max. power (W)
- the test object is allowed to draw as a maximum the
power entered during the function test. If the power is above this value
, a fault P > Pmax is indicated.
Fig. 11.16: Function test parameters - Tripping
Test time (s)
- the test time can be set from 000.1 to 999.9s. The test time
is irrelevant in the case of „Continuous test on“.
Death time „on/off
“ - defines a period before the actual measurement.
The function voltage is applied, the monitoring of current and power is,
however, deactivated for the time entered. This feature is provided for
test objects that drawn high switch on currents, e.g. test objects with
a high capacitive load.
off
- no delay. The current/power evaluation is active.
on
- delay active. The current/power evaluation is switched off.
Death time (s)
- the delay time can be set from 000.1 to 999.9s.