75
10. Menu System
Full-electronic HV-Tester F1-1x Typs
10.9.3 Device
For remote control using PC and communication with an ELABO function
test plug-in or special device.
Test cabinet impuls
Off
- operation without test cabinet and with all other ELABO test
cabinets.
On
- for ELABO 93-4B series test cabinet. The device outputs a pulse
to the test cabinet and as a result opens the lock. The pulse is gener-
ated at the end of the test time or after a „reset“. Is necessary if the
device is used in conjunction with an ELABO 93-4B series test cabinet.
The test time is increased by approx. 250 ms as a result.
Mains frequency
(option) - setting for the measurement circuit (sampling
of the mains frequency). If the setting is incorrect, the measured value
tolerances may be exceeded by a large margin as the sampling rate
does not match the frequency.
50 Hz - the device is operated with a 50 Hz mains frequency
60 Hz - the device is operated with a 60 Hz mains frequency
Serial no. automatic
Off -
The user must enter the serial number, by hand (screen keyboard)
or by barcode scanner. A final "return", will start-up the test run.
ON
- The serial no. is constructed as follow: YYDDD1234567 (year,
day of the year, number). For every test run, the tester will increment
the serial number. For start-up of the test, the user must operate the
"Run"-button, close the test cabinet, etc.
Fig. 10.23: Setting device parameters - Device