Chapter 2. Pump Operations
2.1.6 Preparing the instrument for Pressure/Vacuum operation
Make sure the instrument is safe to use: refer to the instructions on page 27.
Step 1
Power ON the instrument by holding down the power button for
2 seconds until the Druck splash screen appears.
Step 2
Select the required pressure function from the Channel Function
screen.
Note:
From the dashboard select:
CALIBRATOR >>
or
>> FUNCTION >> Pressure>
INT>>Normal
Select the
Tick
softkey and press the
Back
icon.
Step 3
Select the required units.
Note:
From the dashboard select:
CALIBRATOR >>
or
>> UNITS >> either tap the unit
required on the screen or use the Navigation key pad arrow
keys to select the unit.
Select the
Tick
softkey and press the
Back
icon.
CAUTION
To avoid damage to the device under test, hold it firmly while
screwing it into the test port/dirt trap.
Step 4
To attach the device under test to the test port or dirt trap: hold the
device in the socket, then turn the swivel connector at the base of the
connector fully counter-clockwise until it is hand tight.
Note:
Make sure that the device under test has a male G
⅜
Quickfit adapter thread or use a suitable adapter rated to
35 bar.
If in doubt, please contact
www.bakerhughesds.com/druck/global-service-support