222 | DPI610E Instruction Manual–English
Chapter 12. HART
Step 3
Cont’d
Select
OK
in the popup message
window to accept the Trim.
Or select
CANCEL
to return to the
previous step.
Step 4
When the Low trim is complete, select
the
Next
softkey to proceed to the
High Trim procedure or select the
Back
softkey to exit.
Step 5
The second part of the D/A trim
procedure focuses on the High (value)
trim.
The default High Trim value is 20 mA.
This value can be edited if required by
selecting the
Edit
softkey (Step 6)
If no change is required, select the
Next
softkey to proceed (Step 7).
Procedure continues on next page