Copyright 2022 Baker Hughes Company.
English–DPI610E Instruction Manual | 167
Setting up Periodic Data Logging
9.5 Setting up Periodic Data Logging
The
PERIODIC
option is a time-based data trigger (see Section 9 on page 156, Section 9.2 on
page 159 and Section 9.3 on page 160). Use the following steps to continue with a
PERIODIC
triggered Data Log session:
Step 1
After selecting the
PERIODIC
Data Log
mode and the values of the
PERIODIC
options, select the
PLAY
icon in the
Data Log Setup
screen.
The
Calibrator
main screen appears.
Step 2
To start the periodic logging, select the
Periodic
Data
Log
softkey. This is
similar to the icon for the
Key Press
Data
Log
icon except it has a clock in its center
rather than an addition sign.
At this point, the
Data Log Status
icon
appears in the status bar until the logging is
complete. Periodic logging is automatic,
based on the configurations in the
Setup
menu. At the end of the session, a popup
message appears, to confirm that data
logging is complete. The file is saved
automatically.
It is not possible to change the channel
configuration once data logging has begun.
If at any point during data logging, the user
navigates to the Dashboard, some
applications which may interfere with the
logging will be disabled. Such applications
will have a slash across the icon on the
Dashboard.