Chapter 2. Pump Operations
2.1.4 Attach Instrument Dirt (and Water) Trap to Test Port
2.1.5 Attach/Remove Device Under Test
2.1.5.1 Attach Device Under Test
INFORMATION
It is recommended that a dirt trap is always used.
CAUTION
To avoid damage to the dirt trap, hold it firmly while screwing it into
the test port.
To attach the Instrument Dirt (and Water) Trap (IDT) to the test port, first remove
the Blanking Plug if it is in the test socket: turn the swivel connector clockwise to
release the plug. Put the trap into the socket while turning the swivel connector fully
counter-clockwise until it is hand tight.
To avoid damage to the device under test, hold it firmly while screwing it into the
test port/dirt trap.
To attach the device under test to the test port or dirt trap, put the trap into the
thread of the swivel connector, then turn the swivel connector fully counter-
clockwise until it is hand tight.
Note:
Make sure that the device under test has a male G
⅜
Quickfit
adapter thread or use a suitable adapter rated to 35 bar.
If in doubt, please contact
www.bakerhughesds.com/druck/global-
service-support