Copyright 2022 Baker Hughes Company.
English–DPI 610E Instruction Manual | 9
DPI 610E Series
1. Overview
The DPI 610E series instrument comprises a portable pressure calibrator with an in-built
pressure generation manual pump, with pressure measurement and simulation capability. It is
intended to be used to calibrate pressure sensors and transmitters. It has a smart and simple user
interface allowing it to be operated by any technician, service or maintenance engineer.
The DPI 610E series combines a practical and robust design with reliable and accurate
measurements. It is battery-powered and features highly reliable pneumatic and hydraulic
assemblies for continuous and reliable performance for field calibration, even in harsh
environmental conditions.
A flexible hose is supplied with the instrument to enable connection to other equipment.
This calibrator is enhanced with data logging functionality with reliable internal memory for secure
file storage.
The analysis feature includes field error calculations with PASS/FAIL status with the ability to
create or download procedures which can be run on multiple devices, providing calibration
certification for asset management and maintenance purposes.
The DPI 610E supports the HART (Highway Addressable Remote Transducer) communication
protocol and allows basic HART operation and configuration on HART supported devices.
1.1 DPI 610E Series
1.1.1 Firmware Versions
This manual is based on the use of application firmware DK0492 version V01.00.10. For further
details see End User Software Release Note document
169M2231
available at
.
Table 1-1: DPI 610E Series
Photographs of these Calibrator types are on the next page.
Product Name
Case Color
Pressure Range
Calibrator Type
DPI 610E-PC
Blue
0.35 bar - 35 bar
(5 - 500 psi)
(0.035 MPa - 3.5 MPa)
Pneumatic
DPI 610E-SPC
Yellow
0.35 bar - 35 bar
(5 - 500 psi)
(0.035 MPa - 3.5 MPa
Pneumatic - Intrinsically Safe
DPI 610E-HC
Blue
70 bar - 1000 bar
(1000 psi - 15000 psi)
(7 MPa - 100 MPa)
Hydraulic
DPI 610E-SHC
Yellow
70 bar - 1000 bar
(1000 psi - 15000 psi)
(7 MPa - 100 MPa)
Hydraulic- Intrinsically Safe
Overview