104 | DPI610E Instruction Manual–English
Chapter 6. Calibrator Tasks
6.2.10 Voltage
6.2.10.1 Voltage Measure - Configuration:
For the DPI 610E instrument voltage is measured in
Volts (V) or milliVolts (mV) and is a function of
CH2
only.
When the
Voltage
function is selected, the
Direction
is
automatically set to
Measure
, as the Source option is not
supported by the DPI 610E.
The supported voltage range is -30 V to 30 V.
When using the
Voltage
function, there is the additional
option to use the internal 10 V (non-IS units only) or 24 V
power supply provided by DPI 610E or to measurement
of the voltage input from a source that is powered
external to the unit.
Terminal connections for measurement of Volts are
shown in the diagram on the left.
Step 1
From the
Calibrator
screen, select the
channel 2
Setup
screen (either
press the screen icon or press the
softkey.)
Step 2
Select
FUNCTION
from the
CH2
Setup
screen.