232 | DPI610E Instruction Manual–English
Chapter 13. Instrument Calibration
Step 4
Select a
RANGE
option if applicable (e.g.,
20 mA).
Select
to confirm.
Step 5
This screen has three sections.
The top two sections cover:
•
The reference reading from an
external calibrated equipment.
•
The measured (or sourced) reading
from the DPI 610E.
The bottom section provides the following
information:
1.
Function minimum range (or negative
full-scale) value.
2.
Function maximum range (or positive
full-scale) value
3.
Calibration procedure step status
(visual status boxes and text status)
4.
User instructions for each step of the
procedure.
Step 6
To start the calibration, apply nominal
negative full-scale Current (approximately -
20 mA) as per the displayed instruction. This
is the Calibration point 1.
When the measured Current reading is
stable, select the
Next
icon to continue.
Note
:
•
All electrical measure functions
require a 3-point calibration.
•
Current Source function requires a
2-point calibration.
•
Voltage source function requires a
1-point calibration.