Pneumatic System
2.1.5.2 Remove Device Under Test
A rapid drop in pressure can damage the instrument. Open the pressure release
valve slowly and monitor the pressure reading until you reach the desired
pressure
Step 1
Release all of the pressure in the instrument by opening the pressure release
knob fully counter-clockwise.
Step 2
To remove the device under test, hold it firmly while turning the swivel
connector it fully clockwise.
Attach the Blanking Plug to seal and protect the socket if the instrument will
not be in immediate use, see “Use of Blanking Plug” on page 27.