Copyright 2021 Baker Hughes Company.
124 | DPI610E Instruction Manual–English
Chapter 6. Calibrator Tasks
Step 4
Use the keys on the Navigation Pad to
move to the required row and press the
Enter
key to show a screen keypad.
Tap or use the Navigation keys (press
the
Enter
key to enter each number) to
select each keypad number. To confirm
the complete number press the
Tick
softkey.
MEASURED VALUE 1
: a minimum
value within the measurement/source
range of the selected function. By
default, this is populated as the zero or
negative full-scale value of the
measurement/source function.
DISPLAYED VALUE 1
: a minimum
value equivalent to the minimum
measured value expressed as the
custom unit. By default, the value
populated is 0(%).
MEASURED VALUE 2
: a maximum
value within the measurement/source
range of the selected function. By
default, this is populated as the positive
full-scale value of the
measurement/source function.
DISPLAYED VALUE 2
: a maximum
value equivalent to the maximum
measured value expressed as the
custom unit. By default, the value
populated is 100(%).
UNIT LABEL
: this is a free text field
where the custom unit can be named. It
is limited to a maximum of 6 characters.
The default custom unit label is %.
The custom label uses this relationship formula:
Note
: The measured values are in the original units e.g. mA and displayed values are in the
custom label units e.g.%.
When the scaling parameters have been configured, select the
Tick
softkey to confirm the
changes and return to the
PROCESS
screen. The updated
Channel
Scaling
parameters appear
in the
Scaling
fields.
𝐷𝑖𝑠𝑝𝑙𝑎𝑦𝑒𝑑
𝑣𝑎𝑙𝑢𝑒
𝑥
𝐷𝑖𝑠𝑝𝑙𝑎𝑦𝑒𝑑
𝑉𝑎𝑙𝑢𝑒
2
𝐷𝑖𝑠𝑝𝑙𝑎𝑦𝑒𝑑
𝑉𝑎𝑙𝑢𝑒
1
𝑀𝑒𝑎𝑠𝑢𝑟𝑒𝑑
𝑉𝑎𝑙𝑢𝑒
2
𝑀𝑒𝑎𝑠𝑢𝑟𝑒𝑑
𝑉𝑎𝑙𝑢𝑒
1
𝑀𝑒𝑎𝑠𝑢𝑟𝑒𝑑
𝑉𝑎𝑙𝑢𝑒
𝑥